12.13.
X-Pad
12.13.1. X-Pad Calibration
Note: Remove any cup or cover surrounding the tool tip during this
procedure. As a consequence, take care.
In order to work properly, the
X-Pad
itself must be calibrated. The following tools are
applicable when the calibration procedure is carried out:
•
Ballpoint Pen Tool
•
Braille Tool
•
CorruSpeed Knife Tool
•
Drill Tool
•
MultiCUT
•
MultiCUT HP
•
Psaligraphy Knife Tool
•
RM Knife Tool
Procedure
• The calibration wizard is initiated from
Option
->
Table Option
->
Calibrate Measuring Pad
.
• The
Measuring Pad
calibration dialog appears.
• Follow the instructions in the wizard to complete calibration.
Indication that the calibration sequence is active
• A calibration sequence is indicated by:
• Continuous light in the
Laser Pointer
.
•
Start
button light is blinking.