20
5.5 Using the product
Carefully slide the distal end through the endoscope. Exercise no
force.
Flash-overs, activation in the working channel of the endoscope
or without visual control of the distal end of the HF electrode can
cause unwanted thermal injuries to the patient. Therefore, check
that the distal end of the HF electrode protrudes at least 10 mm
out of the endoscope before each activation.
Electrolyte-free solution, e.g. Purisole, can be used for flushing.
20191-333
20191-334
> 4.9 mm
20191-335
20191-336
> 4.4 mm
Instrument
Dimensions of the
instrument tip
(max. projection)
Working channel di-
ameter of the endo-
scope
4,9 mm
4,4 mm
OBJ_DOKU-190022-005.fm Seite 20 Dienstag, 15. Mai 2018 3:59 15
Summary of Contents for 20191-331
Page 2: ...OBJ_DOKU 190022 005 fm Seite 2 Dienstag 15 Mai 2018 3 59 15...
Page 79: ...79 5 5 1 5 2 5 3 1 2 3 OBJ_DOKU 190022 005 fm Seite 79 Dienstag 15 Mai 2018 3 59 15...
Page 87: ...87 8 x X CE OBJ_DOKU 190022 005 fm Seite 87 Dienstag 15 Mai 2018 3 59 15...
Page 88: ...88 OBJ_DOKU 190022 005 fm Seite 88 Dienstag 15 Mai 2018 3 59 15...
Page 179: ...179 6 6 1 2 2 95 C 134 C 6 2 OBJ_DOKU 190022 005 fm Seite 179 Dienstag 15 Mai 2018 3 59 15...
Page 180: ...180 6 3 6 4 1 6 5 pH 5 5 11 1 2 OBJ_DOKU 190022 005 fm Seite 180 Dienstag 15 Mai 2018 3 59 15...
Page 184: ...184 7 8 x X CE OBJ_DOKU 190022 005 fm Seite 184 Dienstag 15 Mai 2018 3 59 15...
Page 207: ...207 8 x CE X CE OBJ_DOKU 190022 005 fm Seite 207 Dienstag 15 Mai 2018 3 59 15...
Page 208: ...208 OBJ_DOKU 190022 005 fm Seite 208 Dienstag 15 Mai 2018 3 59 15...
Page 219: ...219 7 8 batch x X CE OBJ_DOKU 190022 005 fm Seite 219 Dienstag 15 Mai 2018 3 59 15...
Page 220: ...220 OBJ_DOKU 190022 005 fm Seite 220 Dienstag 15 Mai 2018 3 59 15...
Page 221: ...OBJ_DOKU 190022 005 fm Seite 221 Dienstag 15 Mai 2018 3 59 15...