19
5.3 Assembling the product
The HF electrodes can be inserted into the electrode handle in
four different locked positions:
1. Turn the sealing cap on the electrode handle to loosen, but
do not completely unscrew.
2. Insert the HF electrode into the required position as far as it
goes in the electrode handle.
3. Tighten the sealing cap. The HF electrode is fixed into place.
Make sure that the HF electrode cannot be pulled out.
5.4 Endoscope
selection
The HF leakage current at the ocular of the endoscope may
cause thermal injury to the user. In order to prevent this, please
use only endoscopes designed for use with electrosurgical devic-
es.
Select an endoscope with the following working channel diame-
ter:
Instrument
Dimensions of the
instrument tip
(max. projection)
Working channel di-
ameter of the endo-
scope
20191-331
20191-332
> 4.6 mm
4,6 mm
OBJ_DOKU-190022-005.fm Seite 19 Dienstag, 15. Mai 2018 3:59 15
Summary of Contents for 20191-331
Page 2: ...OBJ_DOKU 190022 005 fm Seite 2 Dienstag 15 Mai 2018 3 59 15...
Page 79: ...79 5 5 1 5 2 5 3 1 2 3 OBJ_DOKU 190022 005 fm Seite 79 Dienstag 15 Mai 2018 3 59 15...
Page 87: ...87 8 x X CE OBJ_DOKU 190022 005 fm Seite 87 Dienstag 15 Mai 2018 3 59 15...
Page 88: ...88 OBJ_DOKU 190022 005 fm Seite 88 Dienstag 15 Mai 2018 3 59 15...
Page 179: ...179 6 6 1 2 2 95 C 134 C 6 2 OBJ_DOKU 190022 005 fm Seite 179 Dienstag 15 Mai 2018 3 59 15...
Page 180: ...180 6 3 6 4 1 6 5 pH 5 5 11 1 2 OBJ_DOKU 190022 005 fm Seite 180 Dienstag 15 Mai 2018 3 59 15...
Page 184: ...184 7 8 x X CE OBJ_DOKU 190022 005 fm Seite 184 Dienstag 15 Mai 2018 3 59 15...
Page 207: ...207 8 x CE X CE OBJ_DOKU 190022 005 fm Seite 207 Dienstag 15 Mai 2018 3 59 15...
Page 208: ...208 OBJ_DOKU 190022 005 fm Seite 208 Dienstag 15 Mai 2018 3 59 15...
Page 219: ...219 7 8 batch x X CE OBJ_DOKU 190022 005 fm Seite 219 Dienstag 15 Mai 2018 3 59 15...
Page 220: ...220 OBJ_DOKU 190022 005 fm Seite 220 Dienstag 15 Mai 2018 3 59 15...
Page 221: ...OBJ_DOKU 190022 005 fm Seite 221 Dienstag 15 Mai 2018 3 59 15...