Actual date
Format: DD.MM.YY
Actual time
Format: hh:mm, 24 h display
Summer time change
over mode
Function of the summer/normal time change over.
“Automatic”: Changeover to valid EU recommendations.
“Manual”: Changeover times can be set up in the following addresses.
“OFF”: No changeover.
NT/ST region
In Europe the summer/normal time changeover occurs at a different time to that in the
USA. Select the region in which the unit is to be installed.
Date NT->ST
Day on which changeover from normal to summer time occurs. Format: DD.MM.YY
Time NT->ST
Time at which the clock increases by 1 hour on changeover from normal to summer
time. Format: hh:mm
Date ST->NT
Day on which changeover from summer to normal time occurs. Format: DD.MM.YY
Time ST->NT
Time at which the clock decreased by 1 hour on changeover from summer to normal
time. Format: hh:mm
Release code
Factory default: “0000”, this means changes possible without security code.
Set individual code: Further setting up changes are only possible after entering
this code.
Tip:
Note the code and store out of reach of unauthorised personnel.
Temp. units
Selection of temperature units. All directly connected temperature measurement
using thermocouples or RTDs are displayed in the units selected.
ATA operating mode
•
Barrel memory: Once the ATA flash card is full no further data is stored on the ATA
flash.
Ring memory: The oldest data is always deleted in order to make room for more
up-to-date information.
Pen size normal/wide
•
Determines with which pen size the analogue signal is drawn (normal = 1 dot using
thin pen, wide (fat) = 2 dots with wide pen)
Warning message at xx
min without ATA flash
•
A warning message can be displayed after a preset tme of 1 - 999 minutes after the
ATA flash card has been removed from the unit.
Warning message at
xx %
•
Warns, before diskette is 100 % full. The internal (ring) memory continues to be
used during diskette change or when the diskette is full. The new data is then
written on the new diskette after diskette change (important for complete data
archiving).
5. Set-up unit
72