Memory/Operating mode
Information to memory capacity and selection of Normal/simulation operating modes.
Operating mode
“Normal operation”: The unit operates normally using the actual connected input
signals.
“Simulation”: The unit operates using simulated signals. The active unit settings are
used for this feature. The real measured value display and storage is switched off
during simulation. Instead of this the simulated values are displayed and stored.
Hint:
When returning to normal operation always use the “Signal analysis - Reset”
feature. This avoids falsifying the real values with simulated min, max, and average
values when leaving simulation. If the previous values are required then export them
by saving them on diskette (see chapter “Handling in operation - Available functions -
Diskette”).
Cable open circuit
monitor
Here the channel open circuit monitor (regular voltage pulse) for direct thermocouple
connection can be switched off or on.
Memory info
Information to the size and availability of the internal measured value memory and the
ATA flash memory card or diskette drive. Displays maximum time available for both
when operating on the preset unit settings.
Hint:
Memory info operates under the specified conditions (see “Technical data -
Memory”) of the actual stored unit settings. If changes have been done to the settings
but not yet stored the info will only be updated when set up has been completed and
normal operation resumed (operating “ESC” a number of times). This is also
conditional on “YES” being used to accept the new settings. See “Set-up unit - Set up
- Set-up principle”.
The available memory decreases when:
- Limits/events are monitored and stored
- Digital inputs are used
- Signal analysis is active
- Other groups are to be stored quicker.
5. Set-up unit
89