Fault symptom
Order of troubleshooting
Action or item to be replaced
System unhealthy.
If the pulse is below the laser
threshold, clean the cell mirrors.
Clean the mirrors.
Check the laser beam path for
obstructions.
Clear any obstructions from the laser
beam path.
If the system has rejected
n
consecutive rejects, press the
Reset
alarms
button on LDM.
Press the
Reset
button (see
Status
lamps / errors
).
If the IN/OUT count does not match,
reset the counters.
Reset the counters.
AIR
warning light illuminated.
Check the blower isolator is in the ON
position and the blower status bulb is
illuminated.
Switch the blower isolator ON.
Check the compressed air supply.
•
Locate and turn off the customer
air supply to the leak detector.
•
Examine the connections at the
leak detector and the customer
supply to ensure they are not
damaged.
•
Check the air line is not pinched or
kinked stopping air flow.
•
Turn back on the customer air
supply.
Troubleshooting and diagnostics
Reference Manual
July 2022
00809-0100-4245
90
Rosemount CT4215