Mirror Cleaning Complete
Close and lock the top enclosure door using the
hex allan key.
Note
Cables are connected to the inside of the door.
Care must be taken when closing the door to
prevent damage.
to restart the leak
detection system.
Reset any
indicated on the display panel.
Check the pulse amplitude on the display to
ensure it is > 0.4 V.
Note
if there
are any issues.
7.4
Replace the sample gas inlet filter
The sample gas filter is essential as it prevents dust and debris from reaching the analysis
cell.
Switch of the blower isolator to stop the sample
gas being drawn from the production line
sample head.
Locate the sample gas inlet filter connected to
the rear panel.
A
Circlip retaining washer.
B
Filter housing
C
Replaceable filter module.
D
Sample gas inlet connection
Maintaining the leak detector
Reference Manual
July 2022
00809-0100-4245
82
Rosemount CT4215