A. Mains power connection.
B. Power cable connector (supplied loose)
C. Customer supplied power cable. (minimum required cable
defined by a competent Engineer or technician ensuring cable selection meets
local standards for power demands for the cable length required.
Item
Designation
Type
Rating
Description
A
AC supply
IEC 60309 industrial
right angle connector
2P+E
Yellow: 110 - 130 V
Blue: 200 - 250 V
Fitted to the rear of the leak detector.
B
AC supply
IEC 60309 industrial
cable mount
connector
2P+E
Yellow: 110 - 130 V
Blue: 200 - 250 V
Supplied loose for customer connection.
Pin 1
Live
Pin 2
Neutral
Pin 3
Earth
C
Power cable
Mains 3 core 1.5mm
2
power cable,
sheathed.
16A, 300 - 500V
(minimum)
Customer supplied power cable in
accordance with applicable local and
national codes.
Reference Manual
Installation
00809-0100-4245
July 2022
Reference Manual
39
Summary of Contents for Rosemount CT4215
Page 1: ...Reference Manual 00809 0100 4245 Rev AC July 2022 Rosemount CT4215 Leak Detection System ...
Page 12: ...Contents Reference Manual July 2022 00809 0100 4245 12 Rosemount CT4215 ...
Page 46: ...Start up procedure Reference Manual July 2022 00809 0100 4245 46 Rosemount CT4215 ...
Page 76: ...Software interface description Reference Manual July 2022 00809 0100 4245 76 Rosemount CT4215 ...
Page 86: ...Maintaining the leak detector Reference Manual July 2022 00809 0100 4245 86 Rosemount CT4215 ...