System Ready
Shows when the system is ready for use. This is red during system start
up.
System Healthy
The system can become unhealthy if the gate in/out count is different
or if there have been
n
consecutive rejects (the default value is five
consecutive rejects). Press
Reset counters
on the HMI to clear the
error.
Laser Healthy
The laser becomes unhealthy if it is below the threshold value (see the
red line in ). Check the beam path for obstruction or clean the cell
mirrors as described in . This error automatically clears when the pulse
amplitude is above the threshold.
Maintenance:
The yellow warning triangle appears if the laser pulse amplitude is
below the warning threshold value. System performance is not
affected, but you should consider activating the mirror purge or
cleaning the cell mirrors soon. This error automatically clears when the
pulse amplitude is above the warning threshold.
Too many Consecutive Rejects: If several packs are rejected in a row, this error is triggered.
shows the error message generated. To reset this error, press the
Reset Alarms
button on
the touch screen.
Reference Manual
Software interface description
00809-0100-4245
July 2022
Reference Manual
63
Summary of Contents for Rosemount CT4215
Page 1: ...Reference Manual 00809 0100 4245 Rev AC July 2022 Rosemount CT4215 Leak Detection System ...
Page 12: ...Contents Reference Manual July 2022 00809 0100 4245 12 Rosemount CT4215 ...
Page 46: ...Start up procedure Reference Manual July 2022 00809 0100 4245 46 Rosemount CT4215 ...
Page 76: ...Software interface description Reference Manual July 2022 00809 0100 4245 76 Rosemount CT4215 ...
Page 86: ...Maintaining the leak detector Reference Manual July 2022 00809 0100 4245 86 Rosemount CT4215 ...