Fault symptom
Order of troubleshooting
Action or item to be replaced
Check the 12 and 24V Mini Circuit
Breakers Q1/Q2 have not tripped.
Reset if tripped.
Blower Motor is not switching on.
Check the unit is switched on at the
Blower isolator fitted on the lower side
panel.
If Off switch On the isolator.
Check the MCB fitted to the isolator
distribution box has not tripped.
•
Open the lower enclosure door.
•
If the circuit breaker has tripped.
Switch off the blower isolator and
contact maintenance to
investigate why it has tripped.
Troubleshooting and diagnostics
Reference Manual
July 2022
00809-0100-4245
88