Cell mirrors are precision optical device essential to the continued accurate operation of
the leak detector. Cleaning the mirrors should only be done if purging the cell mirrors does
not the return system laser pulse to normal operating levels, due to the increased risk to
them being damaged during this operation.
Mirrors should only be cleaned by trained, competent people using only the cleaning parts
listed below:
•
Optical grade lens cleaning tissues.
•
Spectroscopic grade methanol or ethanol.
Preparation
Switch off the blower isolator.
Note
Leave the Unit Isolator switched on.
Using a 8mm AF hex allan key give a quarter turn
to open the top enclosure door lock.
Under no circumstance should the lower
enclosure door be opened at this time.
Note
Cables are connected to the inside of the door.
Care must be taken when opening and closing
the door to prevent damage.
With the door open the analysis cell can be
accessed.