A. Product pack.
B. Reject mechanism.
C. Reject chute / bin.
D. Second sample measurement bar.
E. Typical pack handling device.
F. First sample measurement bar.
G. Input gate.
2. Divide the production line speed in m/s (or ft/s) by the maximum pack length in m
or (ft) to determine the time it would take for the pack to pass through the input
gate in msecs (milliseconds). One second = 1000 msecs.
A. Minimum pack length.
B. Maximum pack length.
3. Alternatively, using a stopwatch or video determine the time taken for a pack to
fully pass by the input gate. For fast moving product lines take the measurement
over a larger number of packs and divide the time recorded by the number of packs
Reference Manual