42
Reference Manual
00809-0200-4702, Rev CA
Section 5: Operation and Maintenance: Models 702DX22 and 702DX61
February 2013
Operation and Maintenance: Models 702DX22 and 702DX61
Mapping the PV, SV, and PV_HEALTHY Variables and Parameter
Below is a shot of the gateway screen where the PV, SV and PV_HEALTHY variables and
parameter can be mapped.
Figure 5-10. Smart Wireless Gateway Modbus Register Map
The Fast Fuel Sensor Diagnostics will propagate via the SV variable. This additional
information will provide additional sensor Status information while using the TraceTek
Fast Fuel Sensor.
Warning: If a device becomes not present on the wireless network, Appropriate action
must be taken by the host system.
Important notes regarding the use of Tyco TraceTek Fast Fuel Sensor and TraceTek sensing cable:
Tyco TraceTek sensors must be installed as per manufacturer recommendations.
Do not run the702 transmitter for long periods (more than two weeks) with a Tyco fuel
sensor in the leak state as this will more rapidly deplete the power module.