119
Reference Manual
00809-0200-4702, Rev CA
Appendix B: Product Certifications
February 2013
Product Certifications
IU
ATEX Intrinsic Safety for Zone 2
Certificate Number: Baseefa12ATEX0122X
Applicable Standards: IEC 60079-0: 2011, EN60079-11: 2012
Markings:
3G Ex ic IIC T4 Gc (-60 °C
≤
Ta
≤
+70 °C)
Ex ic IIC T5 Gc (-60°C
≤
Ta
≤
+40°C)
IP66 / IP67
FOR USE WITH EMERSON PROCESS MANAGEMENT SMART POWEROPTION 701PBKKF
Specific Conditions of Use
1.
The surface resistivity of the antenna is greater than 1 G
. To avoid electrostatic charge
build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
2.
The Model 701PB Power Module may be replaced in a hazardous area. The Power
Module has surface resistivity greater than 1 G
and must be properly installed in the
wireless device enclosure. Care must be taken during transportation to and from the
point of installation to prevent electrostatic charge build-up.
IECEx System Certifications
I7
IECEx Intrinsic Safety
Certificate No.: IECEx BAS 07.0082X
Applicable Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T5 Ga (–60 °C
≤
Ta
≤
+40 °C);
Ex ia IIC T4 Ga (–60 °C
≤
Ta
≤
+70 °C)
IP66 / IP67
FOR USE WITH ROSEMOUNT SMARTPOWER ™ POWER MODULE PART NUMBER
753-9220-0001, or FOR USE WITH EMERSON PROCESS MANAGEMENT SMART POWER
OPTION 701PBKKF
WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – SEE INSTRUCTIONS
SWITCH TERMINAL OUTPUT
PARAMETER LIMITS Option Code 32
Switch Terminal Parameters
Option Code 42
Input
Output
U
o
= 6.6 V
U
o
= 6.6 V
U
i
= 26 V
I
o
= 13.4 mA
I
o
= 13.4 mA
I
i
= 100 mA
P
o
= 21.8 mW
P
o
= 21.8 mW
P
i
= 65 W
C
i
= 0.216 uF
Ci= 0.216 uF
Ci= 0.216 uF
Co
IIC
= 23.78 uF
Co
IIC
= 23.78 uF
Li=0
Co
IIB
= 549.78 uF
Co
IIB
= 549.78 uF
Co
IIA
= 1000 uF
Co
IIA
= 1000 uF
Li=0
Li=0
Lo
IIC
= 200 mH
Lo
IIC
= 200 mH
Lo
IIB
= 800 mH
Lo
IIB
= 800 mH
Lo
IIA
= 1000 mH
Lo
IIA
= 1000 mH