A. Isolate (open)
B. Process
C. Equalize (closed)
D. Equalize (closed)
E. Drain vent (closed)
F. Process
G. Isolate (open)
In-line transmitters
Two-valve and block and bleed style manifolds
Isolating the transmitter
In normal operation, the isolate (block) valve between the process port and transmitter
will be open and the test/vent valve will be closed. On a block and bleed style manifold, a
single block valve provides transmitter isolation, and a bleed screw provides drain/vent
capabilities.
A. Transmitter
B. Vent (closed)
C. Isolate
D. Process (open)
Procedure
1. To isolate the transmitter, close the isolate valve.
Reference Manual
Hardware installation
00809-0100-4774
July 2020
Reference Manual
79