Output splitter block
The output splitter (OSPL) function block provides the capability to drive two control
outputs from a single input. It takes the output of one PID or other control block to control
two valves or other actuators.
Index numbers
Table 2-3: Block Index Numbers
Block name
Revision 7
Revision 8
Resource block
1000
1000
Sensor transducer block
1100
1100
Display transducer block
1200
1200
Analog input block
1400, 1500
1400, 1500, 2300, 2400
PID block
1600
1600
Input selector block
1700
1700
Signal characterizer block
1800
1800
Arithmetic block
1900
1900
Integrator block
2000
2000
Control selector block
N/A
2100
Output splitter block
N/A
2200
Function blocks with default block indexes up to 1500 are permanent. Function blocks
with default block addresses 1600 and higher are instantiated and can be deleted.
2.5.2
Modes
The resource, transducer, and all function blocks in the device have modes of operation.
These modes govern the operation of the block. Every block supports both automatic
(AUTO) and out of service (OOS) modes. The blocks may also support other modes.
Changing modes
To change the operating mode, set the MODE_BLK.TARGET to the desired mode. After a
short delay, the parameter MODE_BLK.ACTUAL should reflect the mode change if the
block is operating properly. The automated procedures (methods) make appropriate
resource, transducer, and analog input block mode changes for most configuration tasks.
Permitted modes
You can prevent unauthorized changes to the operating mode of a block. To do this,
configure MODE_BLK.PERMITTED to allow only the desired operating modes. Emerson
recommends always selecting OOS as one of the permitted modes.
Configuration
Reference Manual
July 2020
00809-0100-4774
24
Rosemount 3051
Summary of Contents for Rosemount 3051 CF Series
Page 4: ...4 ...
Page 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Page 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Page 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Page 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Page 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Page 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...