To use guided setup:
a. Navigate to
Configure > Guided Setup
.
b. Select
AI Block Unit Setup
.
Note
Guided setup will automatically go through each step in the proper order.
•
To use manual setup:
a. Navigate to
Configure > Manual Setup > Process Variable
.
b. Select AI Block Unit Setup.
c. Place the AI block in Out of Service mode.
Note
When using manual setup, perform the steps in the order described in
.
Note
For convenience, AI Block 1 is pre-linked to the transmitter primary variable and
should be used for this purpose. AI Block 2 is pre-linked to the transmitter sensor
temperature. You must select the the channel for AI Blocks 3 and 4. The control
host and some asset management hosts can deconfigure the factory assigned links
and assign the primary variable and sensor temperature to other AI blocks.
•
Channel 1 is the primary variable.
•
Channel 2 is the sensor temperature.
If the F
OUNDATION
Fieldbus Diagnostics Option Code D01 is enabled, these
additional channels are available.
•
Channel 12 is the SPM mean.
•
Channel 13 is the SPM standard deviation.
Note
through
are all performed in a single step by step method under
guided setup.
Note
If the L_TYPE selected in
Direct
If the L_TYPE selected is
Indirect
is not needed. Guided setup automatically
skips any unneeded steps.
2. Select the Signal Conditioning
L_TYPE
from the dropdown menu:
a) Select
L_TYPE
: Direct for pressure measurements using the device default
units.
b) Select
L_TYPE
: Indirect for other pressure or level units.
c) Select
L_TYPE
: Indirect Square Root for flow units.
3. Set
XD_SCALE
to the 0% and 100% scale points (the transmitter range):
Configuration
Reference Manual
July 2020
00809-0100-4774
36
Rosemount 3051
Summary of Contents for Rosemount 3051 CF Series
Page 4: ...4 ...
Page 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Page 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Page 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Page 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Page 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Page 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...