Recommended actions
1. Ensure the device is connected to the segment.
2. Check voltage at terminals. It should be 9 - 32 Vdc.
3. Check to ensure the device is drawing current. It should be approximately 17
mA.
Pontential cause
Segment problems
Recommended action
N/A
Pontential cause
Electronics board loose in housing.
Recommended action
Replace electronics.
Potential cause
Incompatible network settings
•
Change host network parameters.
•
Refer to host documentation for procedure.
•
for device network parameter values.
Device does not stay on segment
Note
Wiring and installation 31.25 kbit/s, voltage mode, and wire medium application guide
AG-140 available from F
OUNDATION
™
Fieldbus.
Incorrect signal levels
Refer to host documentation for procedure.
Recommended actions
1. Check for two terminators.
2. Check for excess cable length.
3. Check for bad power supply or conditioner.
Excess noise on segment
Refer to host documentation for procedure.
Recommended actions
1. Check for incorrect grounding.
2. Check for correct shielded wire.
3. Tighten all wiring and shield connections on the effected part of the segment.
Troubleshooting
Reference Manual
July 2020
00809-0100-4774
110
Rosemount 3051
Summary of Contents for Rosemount 3051 CF Series
Page 4: ...4 ...
Page 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Page 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Page 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Page 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Page 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Page 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...