Figure 2-21: Suppressed Device Alerts Screen
A. Screen name
B. Status signal categories
C. Suppress alert check boxes
The screen shown in
is where you can suppress alerts by checking the box next
to the alert to be suppressed.
2.9.7
Alert simulation
Alert simulation provides the capability to simulate configured NE107 or Plantweb
™
alerts.
NE107 alerts and Plantweb alerts show the same consolidated status derived from the
same diagnostics so the single alert simulation is used for both. Alert simulation is typically
used for training or to verify alert configuration.
Menu navigation:
Service Tools
→
Simulate
To enable alert simulation, click the
Enable/Disable Alerts Simulation
button. When
simulate is active it will display on the screen. Once alerts simulation is active, you can
simulate individual alerts by checking the check box next to the desired alert condition.
The device status indication located on the upper right corner of the screen changes to
show the device status associated with the simulated alert. The simulated status is
displayed everywhere device status is displayed. Alert simulation is enabled and disabled
using an automated procedure called a
Method
.
Reference Manual
Configuration
00809-0100-4774
July 2020
Reference Manual
51
Summary of Contents for Rosemount 3051 CF Series
Page 4: ...4 ...
Page 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Page 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Page 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Page 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Page 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Page 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...