2. The status conditions are then mapped into four status signals that can be used for
annunciation on the segment to the host.
3. Before annunciation, a check is made to determine if you have masked any alert
parameters. Any masked parameters will not be annunciated to the host, but will be
visible using the device DD or DTM.
4. Unmasked alert conditions are annunciated by the appropriate status signal to the
host.
Plantweb
™
Alerts and NE107 Alerts are both processed using the steps described above
and annunciate the same consolidated status parameters.
Figure 2-11: NE107 Alert Processing Diagram
1
.
Detailed status includes
conditions found by all
diagnostics the device
runs
.
Detailed status for
NE 107 and PlantWeb
alerts are identical.
2
.
Consolidated status
groups diagnostics by
probable cause and
corrective action
.
Consolidated status for
NE 107 and PlantWeb
alerts are identical.
3
.
Mapping of conditions
defines how conditions will
be reported
.
NE 107
mapping can be user
modified.
4
.
Masking of conditions
determines which
conditions are reported to
the host and which are not
by status signal. All
status signals remain
5
.
Unmasked active
conditions are reported to
the host
.
The unmasked
or PlantWeb Alert
Sensor Status condition 1
Detailed Status
Sensor Status condition N
Electronics Status condition
1
Electronics Status condition
N
Extended Sensor Status
condition
“Sensor Failure”
Extended Electronics
Status condition
“Electronics Failure”
Additional Status
conditions
User Actionable
Consolidated Status
Mapping of Status
Conditions to Status
Signals
FD
_
FAIL
_
MAP
FD
_
MAINT
_
MAP
Additional Status
Signals Mapped
Masking of Alert Parameters
FD
_
MAINT
_
MASK
Alert Conditions reported to host
as NE 107 Status Signals or
Sensor Failure
Electronics Failure
Additional Alert Conditions
by Status Signal
within each status signal
diagnostic conditions and
visible within the device.
conditions are reported by
status signal categories
categories.
PlantWeb Alerts
Configuration
Reference Manual
July 2020
00809-0100-4774
28
Rosemount 3051
Summary of Contents for Rosemount 3051 CF Series
Page 4: ...4 ...
Page 56: ...Configuration Reference Manual July 2020 00809 0100 4774 56 Rosemount 3051 ...
Page 92: ...Electrical installation Reference Manual July 2020 00809 0100 4774 92 Rosemount 3051 ...
Page 102: ...Operation and maintenance Reference Manual July 2020 00809 0100 4774 102 Rosemount 3051 ...
Page 118: ...Troubleshooting Reference Manual July 2020 00809 0100 4774 118 Rosemount 3051 ...
Page 138: ...Reference data Reference Manual July 2020 00809 0100 4774 138 Rosemount 3051 ...
Page 145: ...Reference Manual 00809 0100 4774 July 2020 Reference Manual 145 ...