Configuration and Use Manual
205
Points and the TLP System
P
o
int
s
an
d TL
P
Op
co
de T
a
b
les
S
o
ft P
o
ints
Ca
libration
C.3.3
Parameters
Each point type is associated with a unique set of parameters. For a listing of the parameters available
for each standard point type, see the
ROC Protocol User Manual
.
The parameters associated with the three UDPs are listed below. Note that for UDP 39, parameters
0–50 are displayed in the Select TLP dialog box, but are not used, and are not documented here.
Table C-1
UDP 22 parameters
Parameter
Read/
Write
Data
type
Length
Range
Default
value
Description
0
R/W
UINT8
1
0 , 1
1
Mass Pressure Correction Enable Flag
• 0 = Disabled
• 1 = Enabled
1
R/O
FL
4
NA
0.0
Mass Flow Rate from Sensor
2
R/O
FL
4
NA
0.0
Pressure Corrected Mass Flow Rate
3
R/W
FL
4
NA
–0.0008
Pressure Effect for Mass Correction
(Peffct)
4
R/W
FL
4
NA
30.0
Cal Pressure for Mass Correction (Pcal)
5
R/O
FL
4
NA
0.0
Volume Rate at Flowing Conditions
6
R/W
T_L_P
3
NA
47,0,25
Base Density Input Definition
7
R/W
T_L_P
3
NA
47,0,24
Flowing Density Input Definition
8
R/W
UINT8
1
0 , 1
1
Analog Output Write Enable
9
R/W
T_L_P
3
NA
47,0,0
Source Definition for Analog Output
Value
10
R/O
FL
4
NA
0.0
Value Written to Analog Output
11
R/O
FL
4
NA
0.0
Density Value from Sensor
12
R/O
FL
4
NA
0.0
Density Value from Base Density Input
13
R/O
FL
4
NA
0.0
Density Value from Flowing Density
Input
14
R/O
STRNG20 20
NA
“ ” Density Properties Calculation
Description
15
R/O
FL
4
NA
0.0
Calculation Period (in seconds)
16
R/O
FL
4
NA
0.0
Mass Pressure Compensation Factor
17
R/O
FL
4
NA
0.0
Flowing Pressure Value in PSI
Table C-2
UDP 37 parameters
Parameter
Read/
Write
Data
type
Length
Range
Default
value
Description
0
R/O
FL
4
NA
0.0
Live Data Polling Period
1
R/W
INT16
2
0 – 4
0
Polling State
• 0 = Polling Live Data
• 1 = Reading Setup Data
• 2 = Writing Setup Data
• 4 = Response Timeout
2
R/O
UINT16
2
0 – 65535
245
Live Data Register 1 (Sensor Status)
3
R/O
UINT8
1
1 – 16
FL (7)
Live Data Register 1 Data Type
4
R/O
FL
4
NA
0.0
Live Data Register 1 Value
Summary of Contents for Micro Motion 3711
Page 32: ...24 Micro Motion Model 3711 Gas Flow Computers ...
Page 44: ...36 Micro Motion Model 3711 Gas Flow Computers ...
Page 56: ...48 Micro Motion Model 3711 Gas Flow Computers ...
Page 68: ...60 Micro Motion Model 3711 Gas Flow Computers ...
Page 92: ...84 Micro Motion Model 3711 Gas Flow Computers ...
Page 108: ...100 Micro Motion Model 3711 Gas Flow Computers ...
Page 140: ...132 Micro Motion Model 3711 Gas Flow Computers ...
Page 196: ...188 Micro Motion Model 3711 Gas Flow Computers ...
Page 210: ...202 Micro Motion Model 3711 Gas Flow Computers ...
Page 233: ......