84
Section 7: Troubleshooting
Clean cooling clamp
Personnel: Service technician
Safety gear:
•
Disposable lab gloves
Materials:
•
Fine sandpaper
1. Power down device on rear side using on/off switch.
2. Remove device front panel (
“Remove device front panel” on page 92).
3. Clean cooling clamp (1) on lower cooling system using sandpaper.
4. Install device front panel (
“8.4 Install device front panel” on page 95).
Figure 113 Clean cooling clamp
Replace printer paper (Models 6000P / 6000SP only)
Personnel: User
Safety gear:
•
Disposable lab gloves
Materials:
•
1 Paper roll
1. Power down device on rear side using on/off switch.
1-Paper roll
2-Ribbon
Figure 114 Printer (pulled out)
2. Pull out printer.
Figure 115 Pull out printer