Section 6: Operation
3.
Wait until ice forms on the initiation needle.
After successful ice formation, a calibration prompt or the
Ready to
Measure
screen is displayed (depending on the configured calibration
interval).
Figure 44 Ice formation
4.
If necessary, calibrate device.
.
The device is ready.
Calibration cannot be skipped.
The calibration interval is defined by the lab supervisor.
Figure 45 Calibrate device
5. Tap Measure.
The measurement menu opens.
Figure 46 Device is ready for measurement
6. Pipette the appropriate sample volume into an unused and clean
measurement vessel.
•
Model 6000 / Model 6000P - 15μL
•
Model 6000S / Model 6000SP - 50μL
NOTE!
Risk of incorrect measurement resulting from the presence of air
bubbles in the sample!
• The sample must be pipetted without air bubbles.
Figure 47 Pipette sample