46
Section 5: Setup
Select calibration standards
Specify the allowed calibration points. They should be close to the expected
measurement values.
A calibration point
(Custom)
can be freely selected by the user in increments of 10
and enabled or disabled for selection.
Enter the values in mOsmol/kg.
Figure 28 Select calibration standards
Set up calibration method
4. Select the calibration method:
•
2-point:
Select 2-point calibration
•
3-point:
Select 3-point calibration
5. Select the number of measurements to be performed for each calibration
medium (distilled water and calibration standard).
• The system calculates the calibration values from the mean value of these
measurements.
Figure 29 Set up calibration method
NOTE!
The number of measurements for each calibration point does not affect
the linearity or repeatability of the measurement system.
They are only used to arrive at the mean value.