68
Section 7: Troubleshooting
Error description
Possible Cause
Remedy
Personnel
No crystallization * or
Poor ice crystal
formation in upper
cooling system
Peltier element defective Replace upper cooling system **
Service technician
Cable connection
between cooling system
and PCB defective
Initiation needle too
long or too short
Adjust initiation needle to correct
length (
Initiation needle
not deburred
Debur initiation needle
using fine sandpaper
Initiation needle bent
Manually bend initiation needle
so that it is vertical, replace if
necessary (
Initiation needle
does not move
Initiation needle
disengaged from
motor or stuck
Verify free movement of initiation
needle (
Error in rotation of
motor disk (initiation
needle drive)
Align initiation needle (
Service technician
Correct length of initiation
needle (
Check if light barrier is located
properly and replace if necessary **
Tighten motor disk
Printer does not print
Paper roll is used up
Replace paper roll ()
User
Ribbon is empty
Replace ribbon ()
Measurement procedure
takes longer than usual
Fan is defective
or obstructed
Check if fan is operational and
remove any foreign objects
Replace fan **
Service technician
Peltier element
is defective
Replace lower cooling system **
Spontaneous
crystallization*
Sample not prepared
correctly
Use sample according
to performance data of
device (
User
Thermistor probe not
aligned correctly
Align thermistor probe
(
Service technician
Elevator moved
up by user
Leave the elevator in the
lowered position during the
measurement procedure
User