63
Section 6: Operation
21. Verify the calibration results.
If necessary, tap the individual measurement values to ignore them during
mean value calculation.
( “10.2 Performance parameters” on page 100)
Figure 67 Calibration results
22. Tap Apply to complete the calibration.
The device is now calibrated.
Figure 68 Calibration Results
23.
Verify repeatability (
“Check repeatability of measurements” on page
or
Continue with your measurements (
“Measure individual/batch samples”
).
Note regarding 2-point calibration:
For a 2-point calibration, the measurement accuracy of the device
depends on the correct internal calibration of the thermistor.
If the calibration check reveals above average deviations from the
linearity, the lab supervisor has to recalibrate the internal thermistor:
For this purpose, open the Replace Probe menu (Start menu Lab
Options Maintenance Replace Probe).
Tap Calibrate and perform a 3-point calibration using distilled water
and the two calibration standards with 300 and 850 mOsm/kg.
Figure 69 Replace probe