100
Section 10:
Specifications
Accuracy (Linearity)*
3-point calibration
Models 6000 & 6000P (15µl)
Calibrated Range
mOsmol/kg
0 - 1000 mOsmol/kg
Cal. (0/300/850)
0 - 2000 mOsmol/kg
Cal. (0/850/2000)
0 to ≤ 400
≤ ±4 mOsmol/kg
≤ ±10 mOsmol/kg
> 400 to ≤ 1000
≤ ±1.0%
≤ ±3.0%
> 1000 to ≤ 2000
**
≤ ±1.5%
Models 6000S & 6000SP (50µl)
Calibrated Range
mOsmol/kg
0 - 1000 mOsmol/kg
Cal. (0/300/850)
0 - 2000 mOsmol/kg
Cal. (0/850/2000)
0 to ≤ 400
≤ ±2 mOsmol/kg
≤ ±6 mOsmol/kg
> 400 to ≤ 1000
≤ ±0.5%
≤ ±1.5%
> 1000 to ≤ 2000
**
≤ ±1.0%
*Performance above 2000 mOsmol/kg dependent on selection of calibration values. For best results, it is recommended to calibrate using values
near the intended measuring range. Accuracy (from nominal value, tolerances of reference and calibration solutions excluded) and precision (within
run) specifications apply to ELITechGroup standards and reference solutions.
**Measuring points in this range not recommended for indicated calibration values.