79
Section 7: Troubleshooting
Install elevator cover
Figure 96 Elevator cover with correct cable routing
CAUTION!
Risk of cable break!
20. Place cable (1 Above) on thermistor probe in cable guide (2 Above) and
install elevator cover on device.
21. Bolt elevator cover to elevator using the 2 mounting screws (1 Left).
The thermistor probe is now installed.
Figure 97 Tighten mounting screws
Clean cooling pin
22. Clean cooling pin (1) using a fiberglass pin (AC-200). For this purpose,
move elevator cover up.
23. Power up device on rear side using on/off switch.
24. Calibrate the new thermistor probe.
For this purpose, open the Replace Probe menu (
Start menu Lab
Options Maintenance Replace Probe
).
25. Move elevator cover down.
Figure 98 Clean cooling pin
26. Copy the values from the included data sheet to the fields and tap
Calibrate
.
The system starts the device calibration.
Perform a 3-point calibration on the device (
58). Use distilled water and the two calibration standards with 300 and
850 mOsm/kg.
Figure 99 Calibrate thermistor probe