95
Section 8: Servicing
8.5 Lubricate elevator
Personnel: User
Safety gear:
• Disposable lab gloves
Materials:
• Silicone spray
CAUTION!
Risk of property damage due to wrong lubricant!
The use of lubricating grease can damage the elevator guide.
• Only use silicone spray as lubricant.
1.
Power down device on rear side using on/off switch.
2. Remove device front panel ( “Remove device front panel” on page
Figure 137 Location of elevator guide
3. Lubricate elevator guide on left (A) and right side (B) of ball bearing using
silicone spray.
Tip
The left ball bearing of the elevator guide is not easily accessible.
Therefore, lubricate the upper section (A) on the left side of the
elevator guide and move the elevator up and down multiple times.
Figure 138 Lubricate elevator guide
4. Install device front panel ( “8.4 Install device front panel” on page 94).
Summary of Contents for FreezePoint 6000 Series
Page 1: ...FreezePoint Freezing Point Osmometer USER S MANUAL MODEL 6000 6000P 6000S 6000SP...
Page 2: ......