59
Section 6: Operation
1. Clean thermistor probe using a soft, dry, lint free paper towel.
Figure 53 Clean thermistor probe
2. Follow the instructions on the touchscreen.
Figure 54 Perform zero point calibration
3.
Pipette the appropriate volume of distilled water into an unused and clean
measurement vessel.
•
Model 6000 / Model 6000P - 15μL
•
Model 6000S / Model 6000SP - 50μL
NOTE!
Risk of incorrect measurement resulting from the presence of air
bubbles in the sample!
• The sample must be pipetted without air bubbles.
Figure 55 Pipette distilled water
4.
Position measurement vessel on thermistor probe with cover facing front.
NOTE!
The measurement vessel (1) is securely attached to the thermistor
probe when it clicks into place.
Figure 56 Position measurement vessel