Section 6: Operation
6.4 Measure individual/batch samples
Personnel: User
Safety gear:
• Disposable lab gloves
Materials:
• Soft, lint free paper tissue
• Pipette
• Measurement vessel
• Sample
NOTE!
Impaired repeatability of measurement due to incorrect measurement vessels!
Repeated use of the measurement vessels and use of incorrect consumables cannot guarantee
reproducible measurement results.
• Always use a clean and unused measurement vessel for every measurement.
• Only use measurement vessels supplied by ELITechGroup.
• Never use centrifuge tubes or reaction vessels.
NOTE!
Risk of incorrect measurement resulting from the presence of air bubbles in the sample!
•
The sample must be pipetted without air bubbles.
1. Clean thermistor probe using a soft, dry, lint free paper towel.
Figure 42 Clean thermistor probe
2. Tap Measure on the Start menu.
If the operating time of the device is less than 3 minutes, a wait screen for
ice formation is displayed.
Figure 43 Start menu