Section 7: Troubleshooting
4. Remove initiation needle from motor disk. Hold washer (1) of mounting
screw in place.
NOTE!
Loose washer between initiation needle and motor disk!
Figure 76 Remove initiation needle
5. Remove guide tube (1) and fasteners (3) of initiation needle (2).
Figure 77 Initiation needle components
6. Make sure that the new initiation needle is as straight as possible (1&2).
1-Incorrect orientation
2-Correct orientation
If necessary, bend initiation needle so it is straight (3).
7. Shorten initiation needle to correct length. Proceed as follows.
Figure 78 Straighten initiation needle