Low pressure alarm at elevated compressed air demand.
SIGNALING
– low pressure alarm at
elevated compressed air demand. The
display flashes orange.
PRESSURE - current pressure in the system
The display automatically disappears once
the air pressure is above 3 bar.
The compressor supplies air to the
compressed air system as needed and
without restriction.
Medium priority alarm conditions
Air pump malfunction
The message on the display (FAULT) and
indicator P2-ALARM (17) indicate that the
aggregate is not running (a motor circuit
breaker (Q1 - Q6) is tripped due to current
overload). The display flashes red.
The other aggregates are working normally.
The screen disappears once the malfunction
is remedied and the motor circuit breaker is
manually placed back in the “ON” position.
The display for normal operation is shown.
The compressor only supplies compressed
air to the central line through the functional
aggregates.
All alarms are automatically recorded on an
SD card.
Alarm signals have priority over maintenance interval signals. As such, the light
will indicate an alarm from any of the aggregates.
All error signals are connected to controller output K3:Q3.2 and to terminals
X1:44 and X1:45 (in the control panel) as NON VOLT ALARM SIGNAL.
Data collection
Data is recorded on a micro SD card in a slot
in the base module. Malfunction and
operating event data is recorded on this card.
Information is saved in “.csv” formatted files.
The system sequentially saves data into 50
files, each of which may have up to 20,000
lines.
To copy data from the SD card, manually
remove the card and load the data using
Excel onto a PC or remotely via the Ethernet
network.
Fig. 22: Data collection
Summary of Contents for DK50 4x4VRT/M
Page 1: ...DK50 4x4VRT M DK50 6x4VRT M User manual...
Page 2: ......
Page 4: ...CONTENTS 5...
Page 11: ...PRODUCT DESCRIPTION 03 2021 11 NP DK50 Nx4VRTM AD A EN 4_03 2021 DK50 4x4VRTS M DK 50 6x4VRT M...
Page 15: ...PRODUCT DESCRIPTION 03 2021 15 NP DK50 Nx4VRTM AD A EN 4_03 2021 Fig 1 Compressor with dryer...
Page 36: ...INSTALLATION NP DK50 Nx4VRTM AD A EN 4_03 2021 36 03 2021 DK50 6x4VRT M DK50 6x4VRTS M...
Page 68: ...ANNEX NP DK50 Nx4VRTM AD A EN 4_03 2021 68 03 2021 ANNEX 21 MAPPING PARAMETERS...
Page 69: ...ANNEX 03 2021 69 NP DK50 Nx4VRTM AD A EN 4_03 2021...
Page 71: ......
Page 72: ......
Page 73: ......