Alarms
The equipment has an intelligent monitoring system that generates an alarm
signal based on priority (medium priority alarms have higher priority than low
priority alarms).
Alarm signals have a higher priority than maintenance/service interval signals.
The maintenance/service interval is measured from the first time the equipment is energised. All
alarms are accompanied by a blinking red P2 (Alarm) indicator.
Low priority alarm conditions
Expiry of defined maintenance / service interval
This alarm activates once the 2,000 hour maintenance/service interval expires. The display shows
the following details:
SERVICE
ACCORDING
TO
INSTRUCTIONS FOR USE
TOTAL HOURS - total time the equipment
has been connected to power
RUN HOURS
– equipment operating hours
The display flashes orange.
Note: Press F1 to switch to the normal
operating mode screen for 60 seconds.
The compressor supplies air to the compressed air system as needed and without restriction.
Call in service personnel to perform the required service.
Note: Only service personnel are authorised to configure a new service interval.
Press F4 and hold for at least 5 seconds to confirm the completion of maintenance/service.
The display then changes to the normal operating mode screen.
This indicates the new maintenance interval has been set on the controller from this time forward.
Any maintenance or service work must be recorded in the compressor’s service
log.
Compressor module ambient temperature exceeds the limit.
WARNING - high ambient temperature
alarm.
The display flashes orange.
This alarm appears if the ambient
temperature exceeds the 40°C limit for at
least 30 seconds. The aggregates operate
normally.
The display otherwise shows the current
ambient air temperature.
The alarm clears when the temperature
Summary of Contents for DK50 4x4VRT/M
Page 1: ...DK50 4x4VRT M DK50 6x4VRT M User manual...
Page 2: ......
Page 4: ...CONTENTS 5...
Page 11: ...PRODUCT DESCRIPTION 03 2021 11 NP DK50 Nx4VRTM AD A EN 4_03 2021 DK50 4x4VRTS M DK 50 6x4VRT M...
Page 15: ...PRODUCT DESCRIPTION 03 2021 15 NP DK50 Nx4VRTM AD A EN 4_03 2021 Fig 1 Compressor with dryer...
Page 36: ...INSTALLATION NP DK50 Nx4VRTM AD A EN 4_03 2021 36 03 2021 DK50 6x4VRT M DK50 6x4VRTS M...
Page 68: ...ANNEX NP DK50 Nx4VRTM AD A EN 4_03 2021 68 03 2021 ANNEX 21 MAPPING PARAMETERS...
Page 69: ...ANNEX 03 2021 69 NP DK50 Nx4VRTM AD A EN 4_03 2021...
Page 71: ......
Page 72: ......
Page 73: ......