65
Indication list tab
Located on the right-hand side of the screen is an expandable indications tab. When closed it displays
the number of indications identified above the display threshold. When expanded it lists all indications
for that plate.
Figure 83: Full indication list
When an indication is selected from the list it is highlighted on the scan by the cursor position. The
indication status can be changed to either accepted or rejected using the software dropdown option
or using the dedicated FloormapX tablet button. A short press of the button sets the indication status
to reject, and a long press of the button sets the indication status to accept.
•
Accepted indications can be altered by their surface origin, EPL value or Est. Remain value.
•
Rejected indications are hidden from the scan view but still shown in the list.
Figure 84: Tablet button to change indication status
The indication list button opens the indication list menu which list of all the indications found on every
plate in the inspection.
4.8
Scan Ribbon
In the scan page a graphical representation of the scan captured during acquisition is displayed. The
raw MFL array data is viewed on the left-hand C-Scan and the raw STARS (top surface) data is viewed
on the right-hand C-Scan. Displayed on the far left-hand side is the Endscan which provides a
graphical response from the entire sensor head as the scan is performed. Located on the right-hand
side of the screen is an expandable indications tab. When closed it displays the number of indications
identified above the display threshold. When expanded it lists all indications for that plate.
Figure 85: Scan window during data acquisition
Summary of Contents for Floormap X
Page 1: ...User s Manual Floormap X MFL Array Tank Floor Inspection Solution...
Page 10: ...10 Chapter 1 Precautions and conventions...
Page 16: ...General Precautions and Conventions 16 1 5 EU Declaration of Conformity...
Page 19: ...19 Chapter 2 FloormapX System Overview...
Page 35: ...35 Chapter 3 Inspection Preparation Setup and Operation...
Page 43: ...43 Chapter 4 SIMS GO Data Acquisition Software...
Page 69: ...69 Chapter 5 SIMS PRO...