44)
EBARA
CORPORATION
6.4 Dip Switch
Set the dip switches to select the operating modes as shown in Table 6.5 / 6.6 / 6.7.
Table 6.5 Dip Switch-A Settings
No.
Mode
Off
On
Factory
setting
1
Data Length
7bits
8bits
ON
2
Monitor Cooling water and N2
Always
Only during operation
OFF
3
Buzzer
Not used
Use
ON
4
Operation switched to Remote
According to signal
Automatically stop
OFF
5
External start/stop signal
*
Alternate (Level)
Momentary (Pulse)
*
OFF
6
-----
-----
-----
-----
7
-----
-----
-----
-----
8
BP operation in Remote
Automatic operation
External signal input
OFF
* Optional
Table 6.6 Dip Switch-B Settings
No.
Mode
Off
On
Factory
setting
1
-----
-----
-----
-----
2
-----
-----
-----
-----
3
-----
-----
-----
-----
4
-----
-----
-----
-----
5
-----
-----
-----
-----
6
Remote Interface (IF)
Exclusive special IF
No use / standard IF
ON
7
Phase error monitoring
Standard
During starting only
OFF
8
LCD screen initialize
Carry out initialize
Do not initialize
OFF
Table 6.7 Dip Switch-C Settings
No.
Mode
Off
On
Factory
setting
1
Pump N2 WARNING output
Normal Open
Normal Close
OFF
2
-----
-----
-----
-----
3
Back press WARNING output *
Normal Open
Normal Close
OFF
4
-----
-----
-----
-----
5
-----
-----
-----
-----
6
-----
-----
-----
-----
7
-----
-----
-----
-----
8
-----
-----
-----
-----
* Optional
DIP SW-A. No.1
In case of observing pump running status with RS232C
communication port, Data Length can be selected out of
7bits and 8bits.