ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
45
3.3 Limit table setup display
Press the [Meas Setup] key, and then press the [Limit] function key to enter into the
limit table setup page. The limit table setup page helps users configure the built-in
comparator of ET35.
Limit table setup page
Parameters can be set on the limit table setup page
measurement function (function)
reference nominal value (nominal value)
auxiliary levels (subsidiary)
comparison function limit mode (mode)
comparison function (compare)
lower limit of each level (lower limit)
upper limit of each level (upper limit)
The built-in comparison function can classify the DUT into up to 9 levels (BIN 1 to
BIN8 and BIN OUT). Eight pairs of main parameter limits and a pair of auxiliary parameter
limits can be set. If the main parameter of the DUT is within the limit range of the level and
the secondary outside the range, then the DUT is classified into the auxiliary level. After
HANDLER interface is installed on ET35, the measurement result can be output to the
automatic measurement system to achieve automatic test.