ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
101
completed. /EOM is output after all measurements end for each stepping in the sweep
cycle.
Layout of List sweep comparator pins
PIN No.
Signal name
Description
1
2
3
4
5
6
7
8
9
10
11
/BIN1
/BIN2
/BIN3
/BIN4
/BIN5
/BIN6
/BIN7
/BIN8
/BIN9
/OUT
/AUX
Sweep point 1 exceeds the limit
Sweep point 2 exceeds the limit
Sweep point 3 exceeds the limit
Sweep point 4 exceeds the limit
Sweep point 5 exceeds the limit
Sweep point 6 exceeds the limit
Sweep point 7 exceeds the limit
Sweep point 8 exceeds the limit
Sweep point 9 exceeds the limit
any sweep point exceeds the limit
Sweep point 10 exceeds the limit
12
13
/EXT.TRIG
External trigger: when the triggering mode is set
to EXT.TRIG (external trigger), ET35 is triggered
by the rising edge pulse signal imposed on the
pin.
14
15
/EXT.DCV2
External DC voltage 2: DC supply pin of the
photocoupling
signal
in
the
instrument
(/EXT.TRIG, /KeyLock, /ALARM, /INDEX, /EOM
)
16
17
18
NC
No connection
19
/PHI
Main parameter is too high: the measurement
results are larger than the upper limits of BIN1 to
BIN9.
20
/PLO
Main parameter is too low: the measurement
results are smaller than the lower limits of BIN1
to BIN9.
21
/SREJ
Auxiliary parameter fails: the measurement
results are not within the limits for auxiliary
parameter.
22
23
24
NC
NC
NC