ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
83
Accuracy
±(5%+5mVrms)
Constant
value
Range
10mVrms~1Vrms
Resolution
1mVrms
Accuracy
±(2%+2mVrms)
Current
Rated value
Range
100μArms~20mArms
Resolution
10μArms
Accuracy
±(5%+50μArms)
Constant
value
Range
100μArms~10mArms
Resolution
10μArms
Accuracy
±(2%+20μArms)
(
3
)
Bias voltage
Internal
Range
-2V~2V
Resolution
1mV
Accuracy
±(2%+5mV)
External
Range
-60V~60V
Resolution
Depending on external input
Accuracy
Depending on external input
(4) Level monitoring
Level monitoring includs three items:
VAC: the actual voltage betwen both ends of the DUT
IAC: actual current through the DUT
EBIAS: external bias voltage applied (only displayed in the external bias mode)
VAC
Range
0.0001V~2.000V
Resolution
Four digits and a half
Accuracy
±5%
IAC
Range
0.0001nA~30.00mA
Resolution
Four digits and a half
Accuracy
±5%
EBIAS
Range
±0V~60.00V
Resolution
0.01V
Summary of Contents for ET3501
Page 25: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 22...
Page 77: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 74...
Page 93: ...ET35 Series Digital Bridges LCR Operating Instructions Hangzhou Zhongchuang Electron Co Ltd 90...