ET35 Series Digital Bridges(LCR)Operating Instructions
Hangzhou Zhongchuang Electron Co., Ltd.
84
Accuracy
±
(
5%+3V
)
(
5
)
Output resistance
30Ω, 100Ω optional
6.1.3 Measurement time
Fast, medium, slow, custom.
Slow: 800ms.
Medium: 100ms.
Fast: 20ms for ET3502, ET3503, ET3505, ET3510; 50ms for ET3501.
Custom: 5ms to 2s for ET3502, ET3503, ET3505, ET3510; 20ms to 2s for ET3501.
The measurement time must be an integer multiple of the measurement signal cycle,
so the actual measurement time is related to the frequency of the measurement signal.
The actual measurement time is calculated with the formula:
T
N
real
.
t
N is the number of the measurement signal cycles contained in the set measurement
time (N is 1 when it is actually less than 1), T is a measurement signal cycle.
Mode of
measurement
time
Frequency
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz
Slow
800ms
800ms
800ms
800ms
800ms
800ms
Medium
100ms
100ms
100ms
100ms
100ms
100ms
Fast
100ms
50ms
50ms
50ms
50ms
50ms
Fast*
100ms
20ms
20ms
20ms
20ms
20ms
Note: fast indicates the measurement time of ET3502, ET3503, ET3505, and ET
3510; fast * the measurement time of ET3501.
The actual custom measurement time of ET3502, ET3503, ET3505, and ET3510 is:
Frequency
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz
Upper limit
2000m
s
2000m
s
2000m
s
2000m
s
2000m
s
2000m
s
Lower limit
100ms
10ms
5ms
5ms
5ms
5ms
The actual custom measurement time of ET3501 is:
Frequency
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz
Upper limit
2000m
s
2000m
s
2000m
s
2000m
s
2000m
s
2000m
s
Summary of Contents for ET3501