EOP-D73-001-7-002-05E
E.J. OZONE PRODUCTS
- 16 -
Mass:
Approximately 15 kg
Pipe Port:
(sample inlet (S type,
R type), zero gas inlet
(R type))
Rc 1/4
''
socket
* In the case of connecting the sample gas that can be measured without
problems and has undergone pretreatment (having adjusted the flow
rate and having removed dust and so on), without using the accessory
filter and valve assembly or zero gas and valve assembly (R type).
Pipe ports diameter:
(sample gas inlet)
Select one from the following pipe ports diameter
Ø1/4 inch, Ø3/8 inch, Ø6 mm, Ø10 mm (sample gas inlet)
* In the case of using the accessory filter and valve assembly or zero
gas and valve assembly (R type).
Wire Ports:
Power supply/control cable port: Ø9 mm–Ø22 mm
Signal cable port: Ø4 mm–Ø12 mm
Ambient temperature:
0°C–45°C
Relative humidity:
90 % RH or less within the case (no condensation)
Temperature
compensation:
Measurement range: 0°C–50°C
Temperature compensation: 0°C (Switchable to 20°C compensation)
Others:
Warm-up time is settable for duration of 0 to 99 minutes
(default:10minute)
♦
Option
Pressure compensation:
Compensation range: 80–130 kPa (abs) (optional)
Pressure sensor (Measurement range: 0.35 MPa (abs))
(2) Off-gas Ozone Analyzer
Model:
ME820 Type S
Measuring principle:
UV absorption method
Detection target:
Exhaust ozone gas
Measuring ranges:
Select from the following ranges
0–10.0, 0–20.0, 0–40.0
*unit: g/m
3
(N)
0–50.0, 0–100
0–1000, 2000
*unit: ppm
Measurement cycle:
Continuous measurement
Sampling method:
Suction by internal pump
Materials in contact
with gas:
SUS316, SUS304, PFA, PTFE, PCTFE, PVDF, PVC, Quartz glass,
Alumina, Fluorine-series resin, FKM, Glass fiber, Sapphire, etc.
(Zero gas line)
PVA, FEP, PE, NBR, PBT, POM, HNBR, FKM, SUS304, PVDF,
EPDM, Phenol resin, Felt, Pyrex glass, Glass fiber, Brass, Carbon
steel, Zinc alloy, Nylon, TYGON
®
, etc.
* Ozone destruct: Decomposing catalyst, FKM, Felt