EOP-D73-001-7-002-05E
E.J. OZONE PRODUCTS
- 38 -
(2) Temperature/pressure check (TP)
Temperature (unit: °C) is displayed in sub display 1, and pressure (unit: MPa (G) or hPa (abs))
is displayed in sub display 2.
* If specifications do not include temperature/pressure compensation, actual value is not displayed.
(3) Zero calibration check (ZEro)
Currently selected zero point calibration value is displayed.
(4) Concentration history display (co1)
The past 10 concentration values are displayed in the main display (sub displays 1 and 2).
Press the shift switch to go back to the previous concentration data, and press the increment
switch to proceed to the next concentration data.
* The contents of sub displays 1 and 2 depend on “Temperature/pressure compensation setting.”
(5) Integrated lamp ON time check (LAMP)
Approximate integrated lamp ON time is displayed in unit of 10 hours.
* Perform “integrated lamp ON time clear setting” when the lamp is replaced.
(6) Integrated analyzer usage time check (ToTL)
Approximate integrated analyzer usage time is displayed in unit of 10 hours.
(7) Software version check (vEr)
Software version is displayed.
5.2.4 Test
mode
Press the shift switch and the increment switch simultaneously, and if the [TES] mode indicator
lamp comes on, the “test mode” has been entered. Move to a desired parameter using the shift
switch or the increment switch, and press the enter switch to enter the check (test) item. Press the
mode switch to exit the setting item.
Figure- 9 Typical test mode transition
Enter
Mode
Enter
Test 1: Analog output (TE1)
Test 2: Contact output (1)
(TE2)
Test 3: Contact output (2)
(TE3)
Enter
Mode
Enter
0 00
1 11
1 11
Enter
Mode
Enter
0 00
Test 4: Contact input (TE4)
Enter
Mode
Enter
0.00
0.25
0.50
0.75
1.00