(29) Integrated lamp ON time clear setting (cLr)
Change the display to “on” and press the enter switch, and the integrated elapsed time is
cleared.
See “9.5.1 How to replace the lamp unit” for details.
(30) Pressure sensor atmospheric pressure calibration setting (PAo)
The calibration method varies depending on whether gauge
pressure or absolute pressure is selected in “pressure
compensation type setting”.
See “5.2.2 (13) Pressure compensation type setting” for details.
[1] Gauge pressure calibration (Set to 0.1 MPa (G) or 0.3 MPa (G)).
By selecting “on” and then pressing the enter switch, the pressure at that time is calibrated
to 0 MPa (G).
* If the calibration range is exceeded, “Err” is displayed.
[2] Absolute pressure calibration (Set to 2000 hPa (abs) or 4000 hPa (abs))
The currently selected pressure offset setting is displayed.
Press the shift switch to select the digit to be changed, and press the increment switch to
select a desired value. The digit to be changed blinks.
By holding down the increment switch, the sign is changed between plus and minus.
The settable pressure offset range is ±500 hPa (abs).
* If the pressure offset range is exceeded, “Err” is displayed.
(31) Pressure sensor span setting (PA S)
The currently selected pressure span calibration ratio is displayed.
Select the digit to be changed using the shift switch, and select a
desired value using the increment switch. The digit to be changed
will blink.
The pressure span calibration ratio is settable within the range from 0 to 1.500.
* Perform “pressure sensor atmospheric pressure calibration setting” before performing
“pressure span calibration ratio setting.”
* If the input range is exceeded, “Err” is displayed.
* Find the pressure span calibration ratio by dividing the “actual pressure” by “displayed
value” (sub display 2). (Find the ratio as the setting value “0”.)
See “5.2.3 Check mode, (2) Temperature/pressure check” for details.
Integrated lamp ON time clear setting
(cLr)
Increment
Enter
Mode
oFF
on
Increment
Enter
Pressure sensor atmospheric pressure
calibration setting (PAo)
Enter
Mode
000
Enter
0
00
0
0
0
0
1
0
0
0
0
“ ”
“ ”
“ ”
“ ”
Setting
method
Pressure sensor atmospheric pressure
calibration setting (PAo)
Increment
Enter
Mode
oFF
on
Increment
Enter