EOP-D73-001-7-002-05E
E.J. OZONE PRODUCTS
- 34 -
(24) Zero calibration mode setting (ZEro)
Zero calibration mode can be selected from manual zero calibration, auto zero calibration, and
intermittent (interval) zero calibration.
* Before changing the zero calibration mode, be sure to check the interval setting time and
purge time.
See “5.4 Zero calibration” for details.
(25) Auto zero interval setting (InT1)
The currently selected auto zero interval setting value is
displayed.
The setting range is from 10 to 60 seconds.
Press the shift switch to select the digit to be changed, and
press the increment switch to select a desired value. The digit
to be changed blinks.
* If a value exceeding the input range is entered, “Err” is displayed.
(26) Intermittent zero interval setting (InT2)
The currently selected intermittent zero interval setting value
is displayed.
The setting range is from 0 to 9999 minutes. Press the shift
switch to select the digit to be changed, and press the
increment switch to select a desired value. The digit to be
changed blinks.
* If 0 min. is selected, zero calibration by timer is not performed.
(27) Intermittent zero purge time setting (Pg)
The currently selected intermittent zero purge time setting
value is displayed.
The setting range is from 0 to 99 seconds. Press the shift
switch to select the digit to be changed, and press the
increment switch to select a desired value. The digit to be
changed blinks.
* If select the time setting of 0 sec, zero calibration cannot be
performed. In addition, the zero calibration becomes invalid
when the power is turned on.
(28) Warm-up operation time setting (UP)
The currently selected warm-up operation time is displayed.
The setting range is from 0 to 99 minutes. Press the shift
switch to select the digit to be changed, and press the
increment switch to select a desired value. The digit to be
changed blinks.
* The factory default setting is “10”.
See “7.1 Preparation for measurement (2) Warm-up
operation” for details.
Increment
Enter
Mode
Zero
calibration
mode (ZEro)
MAnU
AuTo
InT
Increment
Enter
0
0
S
0
0
S
0
1
S
0
0
S
“ ”
“ ”
“ ”
“ ”
Setting
method
0
0
00
“ ”
00
0
0
“ ”
00
1
0
“ ”
00
0
0
“ ”
Setting
method
0
0
S
0
0
S
0
1
S
0
0
S
“ ”
“ ”
“ ”
“ ”
Setting
method
0
0
0
0
0
0
0
1
0
0
0
0
“ ”
“ ”
“ ”
“ ”
Setting
method