LCR7000 & RHEOSIGHTS P/N: 974194 |
Rev: A
|
ECO: 52468
www.dynisco.com
From lab to production,
providing a window into the process
-16-
•
Is the 120V power outlet properly grounded? (220 V Europe/Asia)
•
Is the guide bushing in place?
•
Are no objects on the swing arm (eg: beaker)?
•
Is the exhaust hood or snorkel operational?
•
Does the operator have safety glasses?
•
Does the operator have high temp gloves?
•
Is he wearing arm protection long sleeves or lab coat?
•
Has the protective oil been wiped out of barrel?
•
Are the safety shields installed?
The rheometer barrel leaves the factory coated with oil to protect it from rust. This oil
must be removed before accurate rheological data can be obtained. Put two cotton patches
half overlapped directly over the top of the barrel of the still cold rheometer. Use
rheometer-cleaning tool #2 (pictured in the Parts List Section of this manual) to run the
patches up and down the full length of the barrel. The die and die holder nut should always
be removed when cleaning the rheometer barrel. Repeat the process with fresh patches
until the patches come out clean.
Turn the rheometer power switch “ON”. If you hear a continuous beeping sound, turn the
rheometer off and be sure you have connected the load cell correctly, as described under
the Getting Started section. If the rheometer's front panel lights up but there is no
movement on the crosshead, shut the rheometer off, pull out the
EMERGENCY STOP
button, then turn the rheometer back on. Be sure the rheometer safety shield is closed
before attempting to move the rheometer crosshead. If the rheometer cross head moves
up then down and stops it is operating properly and you may continue. If the rheometer is
not working correctly proceed to the trouble shooting section.
At 230º C purge the rheometer with Polyethylene (PE) or Polypropylene (PP). A purge is
simply a charge of material run through the rheometer without collecting data. The purge
run helps eliminate any remnants from previous tests. It also coats and fills micro cavities
in the metals, which can be a cause of variability if the first charge of material is treated like
the following charges.
6.
Before Testing Safety Check
7.
Purging the Rheometer