Perform step 3.1.2.5 –3.1.2.15 for any other load cells that may be present on the sales
order. Before 3.1.2.5 is performed perform 3.1.2.1 –3.1.2.3 then press
until LC#2 is
shown. The next load cell after LC#2 would be LC#3, etc. If a duplicate load cell (a load cell
of the same range) is to be included with the order the LC SEQUENCE option needs to be
indexed alphabetically (A, B, C, etc.). Also, label the load cell sensors with the sequence
character to differentiate them.
Activate main load cell sensor.
Note:
The main load cell sensor is the load cell that is going to be calibrated and tested on
the machine.
Press
SETUP
key on keypad.
Press
four times.
Press
EDIT
smart key.
Press
until load cell sensor of interest is displayed.
Press
SELECT
smart key.
Press
ESC
key on keypad.
R-CAL
Note:
Step 3.1.4 is an exercise to test that the load cell is properly electrically connected.
Press
SYS
on keypad.
Press
TEST
smart key.
Press
once and CH1 TEST DISPLAY should appear at top of display.
Press
SELECT
smart key.
Press
RCAL
smart key.
Auto Identification
Press
SYS
on keypad.
Press
TEST
smart key.
Press
once and CH1 TEST DISPLAY should appear at top of display.