For the Modified-Cross model and Bird-Carreau-Yasuda model, the Edge Device needs to
take some time to find the best fit and this popup window will show up during the process:
More data points mean more processing time.
The interpolation mode is designed to predict rheology test result that is not possible or
not precise or simply not tested with the actual machine. For example, if you want to know
the shear viscosity of the material at a very low shear rate of 0.01 1/s, which is not possible
with the LCR7000, you can use Power Law or Modified Cross model or Bird-Carreau-
Yasuda model to fit the data, then use the formula given to calulate the shear viscosity
according to the shear rate 0.01 1/s.
To apply the interpolation mode, first you need to set the interpolation range and interval.
5.
Interpolation Mode