needs to be indexed alphabetically (A, B, C, etc.). Also, label the PT sensors with the
sequence character to differentiate them.
Activate PT sensor.
Note:
The main pressure transducer sensor is the load cell that is going to be calibrated
and tested on the machine.
Press
SETUP
key on keypad.
Press
five times.
Press
EDIT
smart key.
Press
until PT sensor of interest is displayed.
Press
SELECT
smart key.
Press
ESC
key keypad.
R-CAL
Note:
Step 3.1.4 is an exercise to test that the PT is properly electrically connected.
Press
SYS
on keypad.
Press
TEST
smart key.
Press
twice and CH2 TEST DISPLAY should appear at top of display.
Press
SELECT
smart key.
Press
RCAL
smart key.
Auto Identification
Press
SYS
on keypad.
Press
TEST
smart key.
Press
once and CH1 TEST DISPLAY should appear at top of display.
Press
SELECT
smart key.
Press
ID
smart key.