LCR7000 & RHEOSIGHTS P/N: 974194 |
Rev: A
|
ECO: 52468
www.dynisco.com
From lab to production,
providing a window into the process
-82-
Carefully thread one end (8 pin bendix female connector) of the 7052-56 onto the 8 pin
connector located at left rear (facing from front).
CAUTION:
Do not force connection and
be sure to align all pins
.
Carefully thread the free end (8 pin bendix female connector) of the 7052-56 onto the 8 pin
connector of the mounted PT. Use same caution as mentioned in 3.2.6.1.
Enter device into library.
Press
SYS
on keypad.
Press
DEVICE
smart key.
Press
PT
smart key.
Press
until desired PT# is viewed on display.
Press
NEW
smart key.
Press
until desired PT sensor is viewed on display.
Press
ENTER
smart key.
Press
. PT UNIT is displayed and should read: bar.
Press
. PT CAPACITY is displayed and should read: PT capacity.
Press
. PT SCALING is displayed and should read: 3.33000 mV/V.
Press
. GAIN is displayed and should read: 128.
Press
. PT R-CAL Press
. PT R-CAL is displayed and should read:
84.000%.
Press
. PT SEQUENCE is displayed and should be blank.
Press
ESC
once on keypad.
Press
SAVE
smart key.
Perform step from Section 3 for any additional PTs. If a duplicate PT (a pressure
transducer of the same range) is to be included with the order the PT SEQUENCE option