To Select a Transducer Step by Step
1.
Press the "SETUP" key on the front panel of the rheometer
2.
Scroll Left or Right through the Setup menus using the Arrow Smart Keys.
3.
Locate setup menu "Sen Chanel#1" or "Sen Chanel #2” NOTE: Sensor #1 will always
be the Load Cell on the LCR 7000 and 7001.
4.
When you have located the sensor you wish to change, press the "Edit" Smart Key
on the rheometer display
5.
Using the Arrow Smart Keys scroll Left or Right to locate the proper sensor you wish
to install.
6.
Press the "SELECT" Smart Key to activate the sensor.
The Pressure Transducer or Load Cell you selected is now active.
•
Insures stability prior to Rate Sweep
o
It is often convenient to run a time sweep on a material to assess its tendency to
degrade at the test temperature before collecting data as a function of rate.
•
PVC Example of Time Sweep
PVC is a particularly interesting material to run since, when its stabilizer package fails, the
material cross-links fairly quickly and shows a marked increase in viscosity. Comparing
runs made at different temperatures gives insight into the stable lifetime of a material in an
extruder at various temperatures.
4.
Time Sweep